Allegro Hi-Vol Area Pump

Volumetric Air Sampler (14407) samples air with the smooth, constant flow of a Rotary Carbon Vane. Oil-less operation with only one moving part requires very little maintenance. Regulator has a locked-in operating range of 3-20 lpm. Pump weighs less than 10 lbs. Diaphragm Pump (32451) samples air with lower-cost diaphragm operation. Flow rate: 3-18 lpm. Both pumps include rubber feet for stability, 5 ft. of clear tubing, adjustable aluminum stand and locking regulator. Stand adjusts from 1 ft. to 5 ft. high.





Quest M-28 Dosimeter

Personal Noise Dosimeter:
Accumulates noise dosimetry data.
Accumulates peak and maximum levels.
Accumulates 1-minute histograms or 10 second histograms with peak or maximum levels if de sired.
Stores 100 exceedance levels.
Is housed in a secure dust-proof case.
May be clipped on the belt or worn in the shirt pocket.
Environmental Monitor:
Accumulates up to 120 hours of 1-minute histograms. (If 10-second histograms, peak, or
maximum levels are selected, the total time will be less).
Calculates Ldn levels between 2200 and 0700 hours.
May be tripod-mounted.
The M-28D as an overall unit accumulates data for:
A noise dosimeter.
A 10-second or 1-minute time histogram.
A selectable peak and/or maximum histogram.
An exceedance level distribution.
Sound levels, peak levels, and maximum levels.
Run time, pause time and elapsed or real time.
Day/night levels (Ldn).
The M-28 D displays:
Sound level, peak level, maximum level.
An equivalent sound level (LEQ).
Percent dose (DOSE).
A time weighted average (TWA).
Run time and pause time.
Elapsed or real time clock (12- or 24- hour).
Calibration level.
Code of 24 internal switch settings.
Low battery and overload indicators.
Run, Pause, %, and dB annunciators.

Standards:
ANSI SI .4-1983 type 2
ANSI SI..25-1978 type 2
IEC 651 type 2
IEC 804 type 2

UL Intrinsic Safety:
Class I Group C,D
Class II Group E,F
Class III

Detector:
True RMS; 63 dB pulse range

Display:
Liquid Crystal Display
Range:
.01 to 2999
Annunciators:
OL, %, dB, LOBAT, RUN, and PAUSE

Battery indicator:
From the time of the first low battery indication, 8-hours of battery life remain with the unit on. One week of data retention remains with the unit off.

Calibrator;
External acoustic calibrator
Temperature:
-15° to +50°C operating; -40° to +60°C storage (battery removed)

Humidity:
0 to 95% non-condensing

Magnetic Field:
Negligible below 50 Oersteds at 50 to 60 Hz Effects
Size:
5.5 x 2.8 x 1.4 inches (140 x 70 x 40 mm)

Weight:
14 ounces (400 grams)
Construction:
Cast aluminum housing with tamper-, water-, and dust-resistant security cove




Quest 1800 Sound Level Meter

Standards: ANSI S1.11-1986, Order 3, Type 2, Sub-type C and IEC R225-1966

Center Frequencies: 10 frequencies from 31.5 Hz to 16 kHz.

Frequency Selection: Full manual control or automatic sequential stepping through each frequency.

Power Source: Sound Level Meter.

Size: 3.3 x 2.8 x 1.8 inches (84 x 71 x 47mm)

Weight: 6 ounces (170 grams)





Quest Q-2700 Sound Level Meter with OBF

General: The OB-50 and OB-100 filter is flat within 0.5 dB in rhw passband, with the 3 dB down points at approximately .707 fc and 1.414 fc where fc is the center frequency of the band chosen. The fc/2 and 2fc frequencies are down by approximately 21 dB with the decade points (fc/10 and 10fc) down by greater than 70 dB.

Standards: ANSI S1.11-1986, Order 3, Type 2, Sub-type C and IEC R225-1966.

Center Frequencies:

OB-50: 9 Frequencies from 31.5 Hz to 8 kHz.

OB-100: 10 Frequencies from 31.5 Hz to 16 kHz.

Frequency Selection: Full Manual control (both models) or automatic sequential stepping through each frequency (OB-100).

Power Source: Sound Level Meter.

Size: 3.3 x 2.8 x 1.8 inches (84 x 71 x 47mm)

Weight: 6 ounces (170 grams)





RION NL-11 Type 1 Integrating Sound Level Meter

Applicable Standards: JIS C 1505, IEC 651 Type 1, IEC Draft Type 1

Frequency: Leq, L AE, Lp & Time

Measurement Range by JIS for sound level meter: (A) 35-130 dB, (C) 40-130 dB, (FLAT) 45-130 dB; Internal noise is 8 dB below the minimum measurement level.

Measurement Range by integrating sound level meter standard: (LA eq) 33-145 dB, (LC eq) 33-145 dB, (L FLAT eq) 33-145 dB

Frequency Response: 20-12,500 Hz, 10-20,000 Hz

Microphone: 1/2" Condenser Mic. Model UC-26

Polarization Voltage: DC 60 V

Sensitivity: -32 dB

R.M.S. Accuracy of Detector: (CF•5) 0.2 dB, (CF•10) 0.5 dB

Dynamic Range: 70 dB

Meter Dynamic Chracteristics: FAST, SLOW, IMP, PEAK HOLD

Overload Indication: LED (Instantaneous), LCD (Latch)

Pause Function: Interrupt calculations at any time.

Leq, L AE Calculation Methode, Sampling Time: Digital, Approximately 30 ms.

Measurement Time: 10 s, 1 min, 5 min, 10 min, 1 h, 8 h, 24 h, Man (Max. 100 h)

Display Analog Meter: Scale Range: 30 dB; Range:70 dB, 3 Step

Display Digital Meter: Indication Cycle: 1s; Resolution: 0.1 dB; Indication: Lp, Leq, LAETime

Calibration: Electrical calibration by internal oscillator (1000 Hz, Sin. Wave)

Power: AA-size ( UM-3, AM-3 or R6, LR6 ) Battery x 4 (20 h of continuous use with alkaline batteries or 12 h with manganese batteries) AC Adapter NC-34

Ambient Condition of Operation: -10 --50°C, 10-- 90% RH

Dimensions: Approximately 26 x 11 x 6.5 cm

Weight: Approximately 850 grams







Alnor Balometer

Accuracy: ±3% of full scale, except
±20 CFM on 250 CFM scale
(±35 cmh on 400 cmh scale)
(±10 l/s on 120 l/s scale)

Operational Temperature Range: 32–122°F (0–50°C)

Storage Temperature -40 to 140°F (-40 to 60°C)

Supply and Exhaust Ranges: 250, 500, 1000, 2000; Standard CFM (note 1). (400, 800, 1700, 3400 standard cmh). (120, 240, 475, 950 standard l/s)

Scale Divisions: 5 CFM from 25 to 200 CFM, 10 CFM from 100 to 500 CFM, 20 CFM from 400 to 1000 CFM, 50 CFM from 800 to 2000 CFM , 10 cmh from 50 to 350 cmh, 10 cmh from 200 to 850 cmh, 25 cmh from 700 to 1700 cmh, 50 cmh from 1400 to 3400 cmh, 5 l/s from 10 to 100 l/s, 5 l/s from 50 to 240 l/s, 10 l/s from 200 to 475 l/s, 25 l/s from 400 to 950 l/s

Maximum Usable Limit: 2000 CFM (3400 cmh, 960 l/s)

Read-Out Time: Approx. 4 seconds

Standard Opening: 2 x 2, 2 x 4, 1 x 4, 1 x 5, 3 x 3 ft (60 x 60, 60 x 120, 30 x 120, 30 x 150, 90 x 90 cm)

Dimensions: Instructions: Height 40 inches (102 cm) Width, Depth --variably depending on the cloth hood size. Up to 5 ft. (153 cm) wide, 3 ft. (92 cm deep at top opening.) Base 17” x 17” (43 cm x 43 cm); Carrying Case: 13” x 26” x 23” (H x W x D) (33 x 66 x 59 cm)





MSP 400

Flow rate: 10 L/min
D50 Diameter: 10 µm or 2.5 µm
Filter Diameter: 37 mm
Dimensions: (L x W x H) 240 x 230 x 300 mm (9.4" x 9.1" x 11.8")
Weight: 3.6 kg (7.9 lb)
Power: 115 VAC, 60 Hz, 5A




Andersen N6 Bioaerosol Kit

Flow rate
Standard BioStage: 28.3 L/min
BioStage 200: 14.15 L/min
Material
Inlet cone and base plate: Precision-tooled autoclavable aluminum
O-rings: Duro 50, BUNA-N
Jet Classification Stage
Standard: 400 holes
BioStage 200: 200 holes
Sample Media
90 to 100-mm agar plates*
Suggested media
For bacteria:
Tryptic Soy Agar (TSA)
Blood Agar Plates (BAP)

For fungi:
Potato Dextrose Agar (PDA)
Malt Extract Agar (MEA)
Dichloran Glycerol 18 Agar (DG-18)
Corn Meal Agar (CMA)
Analysis
Colony culture




8 Stage Viable Impactor Anderson Sampler

Standard Flow Configuration

Operating at 28.3 lpm (um)

Stage 0 = 9.0

Stage 1 = 5.8

Stage 2 = 4.7

Stage 3 = 3.3

Stage 4 = 2.1

Stage 5 = 1.1

Stage 6 = 0.7

Stage 7 = 0.4

60 lpm Flow Rate Configuration Kit Installed

Operating at 60 lpm (um)

Stage -1 = 8.6

Stage -0 = 6.5

Stage 1 = 4.4

Stage 2 = 3.3

Stage 3 = 2.0

Stage 4 = 1.1

Stage 5 = 0.54

Stage 6 = 0.25

90 lpm Flow Rate Configuration Kit installed

Operating at 90 lpm (um)

Stage -2 = 8.0

Stage -1 = 6.5

Stage -0 = 5.2

Stage 1 = 3.5

Stage 2 = 2.6

Stage 3 = 1.7

Stage 4 = 1.0

Stage 5 = 0.43




Sling Psychrometer

225-570: -20° to +120°F (1° graduations)
225-571: -30° to +50°C (0.5° graduations)
Size: 12" L x 1.5" W (305 mm x 38 mm)
Weight/shipping: 0.4 lb/1 lb (0.2 kg/0.5 kg)




TSI Q-Trak 8550


CO2:
Sensor type:............................. Non-Dispersive Infrared (NDIR)
Range:...................................... 0-5000 ppm
Accuracy:................................. ±3% of reading ±50 ppm at 25°C
(77°F). [Add uncertainty of
±0.36% per °C (±0.2% of reading per
°F) for change in temperature.]
Resolution:............................... 1 ppm
Response time:........................ 20 seconds (for 63% of final value for
500 ppm step change)

Temperature:
Sensor type:............................. Thermistor
Range:...................................... 0 to 50 °C (32 to 122 °F)
Accuracy:................................. ±0.6 °C (1.0 °F)
Resolution:............................... 0.1 °C (0.1 °F)
Response time:........................ 30 seconds (90% of final value, air
velocity at 2 m/s)
Display units:............................ °C or °F (user selectable)

Humidity:
Sensor type:............................. Thin-film capacitive
Range:...................................... 5 to 95 %rh
Accuracy:................................. ±3 %rh (includes ±1% hysteresis.)
Resolution:............................... 0.1 %rh
Response time: ....................... 20 seconds (for 63% of final value)

Instrument Temperature Range:
Operating range:...................... 5 to 45 °C (41 to 113 °F)
Storage range:......................... -20 to 60 °C (-4 to 140 °F)

Time Constant:
Range:...................................... Adjustable from 2 to 60 seconds

Data Logging:
Data points:.............................. 63,000 (14 days of logging simultaneous CO2, Temperature, and Humidity at 1-minute intervals).
Logging interval:....................... User adjustable from 1 second to 1 hour

Power Requirements:
Batteries:.................................. Four AA-size Alkaline or NiCd
or
AC adapter:.............................. 6 VDC nominal, 300 ma, [Q-TRAK
mates with 5.5 mm OD x 2.1 mm ID
plug, center pin positive(+)]
Approx. battery Life:................. 13.5 hours (Alkaline), 4.75 hours
(NiCd)

Physical:
External dimensions:................ 107 mm x 183 mm x 38 mm
(4.2 in x 7.2 in x 1.5 in)
Probe length:............................ Model 8551: 17.8 cm (7.0 in)
Model 8550: 14.7 cm (5.8 in)
Probe diameter:........................ 1.9 cm (0.75 in)
Weight:..................................... 0.59 kg (1.3 pounds) (with batteries)
Display:..................................... 4-digit LCD, 15 mm (0.6 in) digit
height

Maintenance Schedule:
Factory calibration:................... Annually
User calibration:....................... As needed

Serial Interface:
Type: ....................................... RS-232
BAUD rate: .............................. 1200
Data bits:.................................. 8
Stop bits:.................................. 1
Handshaking:........................... None
Data format:.............................. ASCII

Calibration Gas
CO2 zero:................................. Nitrogen or pure air (<10 ppm CO2)
CO2 span:................................. 1000–5000 ppm CO2
CO zero:................................... Pure air (<0.5 ppm CO)
CO span:.................................. 35–500 ppm CO
Gas flow rate:........................... 0.3 to 0.5 L/min




TSI Indoor Air Quality Cal 8760


Room Pressure Controller :


Room Pressure Display
Range..................................-0.20000 to +0.20000 inches H2O
Resolution..........................5% of reading
Display Update...................0.5 sec

Inputs
Setback Input......................SPST (N.O.) Switch. Closing switch
initiates condition.
Emergency Input................ SPST (N.O.) Switch. Closing switch
initiates condition.

Outputs
Low Alarm Range..............-0.20000 to +0.20000 inches H2O
High Alarm Range.............-0.20000 to +0.20000 inches H2O
No Flow Alarm Range.......-0.20000 to +0.20000 inches H2O
Alarm Contacts..................SPST (N.C.) 0.5 amps at 10 watts max., 24
VAC max. Contacts close in alarm
conditions and loss of power.
RS-232...............................Yes
RS-485...............................Yes

Analog Output
Type....................................0 to 10 VDC or 4 to 20 mA
Range..................................Variable -0.100 to +0.100 inches H2O -0.0100 to +0.0100 inches H2O -0.00100 to +0.00100 inches
H2O
Resolution..........................1.0% of range for 0-10 VDC
1.2% of range for 4-20 mA


Operating Temperature.............32 to 120°F
Input Power................................24 VAC, 6 watts max
Dimensions................................5.5 in. x 3.125 in. x 1.125 in.
Weight........................................0.75 lb.

Pressure Sensor
Dimensions (DxH)............2.75 in. x 1.25 in.
Weight................................0.2 lb.
Temperature Compensated
Range..................................55 to 95°F
Power Dissipation.............0.16 watts at 0 inches H2O, 0.20 watts at
0.00088 inches H2O
Cable Specifications.........4-conductor, 22 AWG, 4-pin polarized at
both ends with a standard length of 15 ft

Transformer
Primary...............................120 VAC
Secondary...........................24 VAC
Rating..................................20 VA, 0.5 amps max
Dimensions (HWD)..........2.0 in x 2.25 in. x 1.5 in. (mounts to 4 in. x
4 in. duplex box)
Weight................................1.3 lb.
UL/CSA Approved.............Yes
Cable Specifications.........2-conductor, 22 AWG with standard length
of 25 ft., maximum length of 100 ft.

Damper/Actuator
Standard Materials.............16-gauge galvanized, Type 304 SS, or
Type 316 SS.
Optional Coating................Plasite No. 7122 finish
Types of Actuators............Electric or pneumatic
Input Power........................Electric: 24 VAC, 7.5 watts max.
Pneumatic: 24 VAC, 3 watts max.
Time for 90° Rotation.......5 seconds




TSI Portacount Plus

Size: 9.5 inches x 7.5 inches x 5.5 inches (24 cm x 19 cm 14 cm)

Weight: 3.3 lb (1.5 kg)

Fit Factor Range: 1 to greater than 10,000

Concentration Range: 0.01 to 5 x 105 particles/cm3

Particle Size Range: 0.02 to greater than 1 micrometer

Typical Fit Factor Accuracy: +/- 10% of reading up to fit factors of 10,000

Temperature Range: Operation: 32 to 100°F (0 to 38°C) Storage: -40 to 160°F (-40 to 70°C)

Flow Rate: Sample: 100 cm3 /min; Total: 700 cm3 min(nominal)

Power Requirements: 5 VDC, 1 amp peak

Alcohol: Hours per charge: 8 hours at 70°F (21°C); Alcohol Type: Reagent grade isopropyl (99.5% or better)

RS-232 Output: Baud rate: User-Selectable: 300, 600, 1200, 2400, 9600 (factory set at 1200)

Carrying Case: Size: 21 inches x 14 inches x 8.3 inches (53 cm x 36 cm x 21 cm) Weight with Standard Accessories: 13 lb (5.9 kg)

Pass/Fail Settings: User Selectable: 0, 100, 250, 500, 1000, 1250, 1667, 2000, 4000, 5000, 6667, 10000 (other values can be programmed by user)

Factory Recalibration Interval: 1 year




Sensidyne Colorimetric

In order to provide a back up for the on-line analyzers, colorimetric indicator tubes
(National Draeger, Inc., Pitt._.burgh,Pennsylvania) were used for the selected gas
measurements. Irl add!tion to the tube measurements indicated in Fig. 1,tubes for
H_Sand SO2were employed for certain tests. These tubes are normally used with
lt a hand pump for atmospheric sampling. They may, however, be advantageously
used with personal sampling pumps where the sampling rate and duration are
carefully measured. The six tubes were connected to individually
calibrated sampling pumps (DuPont Alpha 2 and DuPont P-4000, DuPont
Company Instrument Systems, Kennett Square, Pennsylvania) and operated for
exactly ten minutes at sampling rate: of 1O0 mL/minute or 50 mL/minute,
depending on the range of each individual tube. The Draeger tube_ were chesen
because of their low pressure drop at the specified flow rate. Their specificity for
the target compound is typically very good, and possible interferences are well
documented by the manufacturer. For ease of use and deployment, ali the Draeger tubes with their associated pumps
as well as the triple sorbent traps and pumps were mounted on a 3 ft x 6 ft board.
A stainless steel tube manifold with quick connect tube fittings was mounted at the
top of the board. Rapid connection of the unit to the overall sampling train as
well as change of individual tubes and pumps was facilitated. Flow rates through
the Draeger tubes were carefully chect:ed prior to each test run using a volumetric
standard (Buck model M-5 primary gas flow calibrator, A.P. Buck Inc., Orlando,
FL.) .

Pump calibration was crucial to the accuracy of the measurements.
Careful volumetric determinations were made prior to and subsequent to each test or run.
Volumes were checked using a Buck Model M-5 Primary Gas Flow Calibration
(A. P. Buck, Inc., Orlando, Florida). During operation, ali pumps were started
immediately prior to a test firing and were allowed to run exactly ten minutes.
The pump flow rate settings were such that the volume of air drawn through each
tube made the tube direct reading in average concentration over a 10 minute
sampling time. That is, if the stain reached the 10 ppm line over the course of the
10 minute sampling period, that meant that the concentration in the chamber
averaged 10 ppm during that time period.




MSA Kwik-Draw

Detects a variety of gases and vapors with one hand operation. Automatic stroke counter eliminates the potential for human error. Easy-to-read stroke indicator "blinks" after proper amount of air has been drawn for successful sampling. Comfortable hand grip can be used even with gloves. Includes remote sampling adapter and handy carrying pouch. 10' sampling line and detector tube handbook sold separately. Interchangeable with Draeger tubes and pumps.




Waterra WS-1 Water Level Tapes

Weight: 4.5 lbs

Dimensions: 12" H x 11.5" W x 3.25" D

Tape Width: 3/8"

Tape Material: Polyethylene coated

Tape length: 100' or 30 meters




HNU PID DL-101

Specifications

Certification: designed to meet Class I, Division II, Groups ABCD
Detection Principle: photoionization
*Range: 0.1 to 2000 ppm
*Linear Range: 0.1 to 400 ppm
*Detection Limit: 0.1 ppm
*Repeatability: +/-l%
Response Time: <3 seconds to 90% response
Inlet flow: 225 cc/min
Ambient Humidity: to 95% RH (noncondensing)
Ambient Operating Temperature: to 40° C (temperature compensated so that a 20° C change in temperature corresponds in reading of <+/-2% full scale at maximum sensitivity)
Calibration: benzene referred default plus 11 user defined calibrations
Interchangeable lamps: 11.7, 10.2, and 9.5 eV
Dimensions: probe (8 in. x 3 in.)
readout module (8 x 3 x 6 in.)
cable (36 in.)
Weight:
probe (3 lbs.)
readout module (4 lbs.)
total weight (7 lbs.)

Operating Time on BatteryContinuous Use: >eight hours continuous use at 23° C
Approximately 6.4 hours at 0° C.
Recharge Time from full discharge: 12 to 14 hours
Recorder Output: 2 V +/-3% at 2000 ppm
NOTE * When equipped with 10.2eV lamp and measuring benzene. Values vary for other compounds, conditions, and lamp used. Use of multipoint calibration can increase the linear range.